The
HI96822 is a rugged, portable digital refractometer designed for the
measurement of salinity of natural or artificial seawater. The HI96822
displays results in three popular measurement units: Practical Salinity
Units (PSU), salinity in parts per thousands (ppt) and specific gravity
(S.G. (20/20)). The instrument's high accuracy and simple operation
gives reliable results each and every time. All readings are
automatically compensated for temperature variations and displayed
within a 1.5 second response time. The sealed flint glass prism and
stainless steel well are easy to clean. Just wipe with a soft cloth in
preparation for the next sample.
- Sample size as small as 2 metric drops (100 μl)
- Sealed stainless steel well with high-grade optical prism made of flint glass
- Fast 1.5 second response time for temperature compensated readings
The HI96822 Digital Refractometer eliminates the
uncertainty associated with mechanical refractometers by measuring the
refractive index to determine the salinity of natural and artificial
seawater including brackish, ocean water or saltwater aquarium. The
instrument converts the refractive index of a sample to PSU, ppt and
specific gravity. These conversions are based on respected scientific
publications using the physical properties of seawater (not sodium
chloride). The HI96822 is an easy to use tool for measuring salinity in
natural and artificial seawater whether in the field, lab or at home.
- Designed for seawater salinity analysis (not NaCl)
- Automatic Temperature Compensation
- High accuracy measurements displayed as PSU, ppt and specific gravity
Making a Standard Solution
Sodium Chloride solutions can be used to check the accuracy of the
meter. The table below lists two Sodium Chloride solutions and their
expected ppt Seawater value.
To make a Standard NaCl Solution (g/100 g), follow the procedure below
- Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.
- Tare the balance.
- To make an X NaCl solution weigh out X grams of high purity dried
Sodium Chloride (CAS #: 7647-14-5: MW 58.44) directly into the
container.
- Add distilled or deionized water to the beaker so the total weight of the solution is 100g.
Example Standard NaCl Solution
NaCl (g)
Water (g)
Total
Expected Seawater value (ppt)
3.5% NaCl
3.50
96.50
100.000
34
10% NaCl
10.00
90.00
100.000
96