Your accuracy and speed of your results are dependent on your pH
sensing bulb and ORP platinum tip being clean. pH and ORP electrodes
also need a free flowing junction to function.
Milwaukee's cleaning solution helps both keep sensing parts clean and
the junction open. Simply open the bottle, fill a beaker or container
so that the level covers the sensor and junction, and soak for about 15
minutes followed by a soaking in MA9015 Storage Solution for at least 2 hours or overnight if you have more time.
In general, and depending on the frequency of use and the
type of application, soak your pH or ORP electrode once per week in
MA9016 Cleaning Solution for about 15 minutes. After soaking, remove
the electrode, rinse with Distilled Water and put the MA9015 Storage
Solution in the probe cap before storage. This will keep your pH or
ORP sensor hydrated and ready for the next use.